Product Description
Up for sale is a Dash 3000 Patient Monitor manufactured by GE Medical.
- Item is in used condition, has been tested and does power on. The plastic is starting to crack along the top and sides of item. Monitor is missing the top piece of handle.
-
This item may be a floor model or store return that has been used, and may not be exact as model pictured.
GE Dash 3000 Features
The GE Dash 3000 is a portable patient monitor. Its lightweight, durable design makes it perfect for both transport and bedside monitoring. The Dash 3000 comes with options for wired and wireless networking to the facility's computer systems. The Dash patient monitor can monitor many different parameters at the same time, some of these include 3 to 5 lead ECG, Pulse Oximetry, non-invasive blood pressure, temperature, and cardiac output. The Dash 3000 has an 8.4-inch color LCD screen where you can view 3 or 6 waveforms and battery capacity on the screen all at once.
The GE Dash 3000 comes with two batteries that can provide up to 5 hours of running time and can be recharged within 2 to 4 hours. The batteries can be charged one at a time to lose any functionality of the monitor while charging.
- Networking options for both wired and wireless
- 3 to 6 waveforms and up to 4 pressure can be viewed at once
- 4-5 hours of running battery time
- Multi-lead ECG analysis and DINAMAP NIBP technology all in one monitor
- seamless Patient Data transfer to Facilities computer system
Alarms
- Categories: patient status and system status.
- Priority: 4-levels crisis, warning, advisory, message.
- Notification: audible and visual.
- Settings: default or user-defined.
- Silencing: 1 minute, current alarm only.
- Pause: 5 minutes in adult ICU mode, 3 minutes in Neonatal ICU mode, and 5, 15, minutes or permanent pause in OR mode.
ECG
- Standard leads available: I, II, III, V, aVR, aVL, and aVF.
- Leads analyzed simultaneously: I, II, III, and V (multi-lead).
- Lead fail: identifies failed lead.
- Alarms: user-selectable upper and lower heart rate limits.
Respiration
- Measurement technique: impedance variation detection. Range respiration rate: 0-200 breaths per minute. User-selectable upper and lower respiration rate limits, and user-selectable apnea limit.
Temperature
- Number of channels: 2.
- Resolution: ±0.1°C.
- Alarms: user-selectable upper and lower limits for T1, T2.
Invasive Blood Pressure
- Number of channels: 2.
- Range: -25mmHg to 300mmHg.
- Alarms: user-selectable upper and lower limits for systolic, diastolic, and mean pressures.
NIBP
- Measurement technique: oscillometric.
- Displayed parameters: systolic, diastolic, and mean pressures, time of the last measurement.
- Measurement modes: manual, auto, and stat in Adult ICU and OR modes, manual and auto in neonatal mode.
- Systolic range: 30-275 mmHg adult, 30-235 mmHg pediatric, 30-135 mmHg neonate.
- Diastolic range: 10-220mmHg adult, 10-220mmHg pediatric, 10-110 mmHg neonate.
Pulse Oximetry
- Parameters monitored: SpO2 and peripheral pulse rate.
- SpO2 range: 50-100%.
- PPR range: 30-300 beats per minute.
- SpO2 accuracy: ±2%, 70 – 100% SpO2, ±3% (50-69% SpO2).
- PPR Accuracy: ±3 beats per minute.
MEDICAL DISCLAIMER
"The sale of this item may be subject to regulation by the U.S. food and drug administration and state and local regulatory agencies. If so, you must be an authorized purchaser. If the item is subject to FDA regulation, I will verify your status as an authorized purchaser of this item before shipping of the item. Regardless of the origin of this equipment, documentation provided or identification appearing upon the equipment, the equipment described and offered here is in no way certified for, recommended for, or offered for any specific use. The purchaser agrees that the seller shall not be held responsible or liable for any injuries or damages, whether incidental or coincidental, associated in any way with the equipment. The purchaser, by bidding or buying this equipment, indicates their knowledge of, and agreement to the terms of this disclaimer."